Tektronix TDS3012C, TDS3014C, TDS3032C, TDS3034C, TDS3052C, TDS3054C Digital Phosphor Oscilloscopes

 


Tektronix TDS3000C Digital Phosphor Oscilloscopes

TDS3000C

Buy It Now !  Ordering Info

TDS3012C

TDS3014C

TDS3032C

TDS3034C

TDS3052C

TDS3054C

TDS3000C Series Characteristics 

Model
TDS3012C TDS3032C TDS3052C TDS3014C TDS3034C TDS3054C
Bandwidth 100 MHz 300 MHz 500 MHz 100 MHz 300 MHz 500 MHz
Channels 2 2 2 4 4 4
Sample Rate
(Ea Channel)
1.25 GS/s 2.5 GS/s 5 GS/s 1.25 GS/s 2.5 GS/s
5 GS/s
Maximum Record Length 10 K points on all models
Vertical Resolution 9-Bits on all models
Vertical Sensitivity (/div) 1 mV - 10 V on all models
Vertical Accuracy ±2% on all models*1
Max Input Voltage () 150 VRMS CAT I on all models (300 V CAT II with standard 10X probe)
Position Range ± 5 div on all models
BW Limit 20 MHz 20, 150 MHz 20, 150 MHz 20 MHz

20, 150 MHz

20, 150 MHz
Input Coupling AC, DC, GND on all models
Input Impedance Selections

1 MΩ in parallel with 13 pF or 50 Ω on all models

Built-In Ethernet
Yes, all models
Timebase
Range (/div)
4 ns - 10 s/div 2 ns - 10 s/div 1 ns - 10 s/div 4 ns - 10 s/div 2 ns - 10 s/div 1 ns - 10 s/div
Accuracy 200 ppm
Display Monitor Color LCD
*1 = Derated at 0.025%/°C for temperatures above +30°C and below +18°C.

Acquisition Modes

DPO - Captures and displays complex waveforms, random events and subtle patterns in actual signal behavior. DPOs are able to provide 3 dimensions of signal information in real-time: amplitude, time and the distribution of amplitude over time.
Peak Detect - High frequency and random glitch capture. Captures glitches as narrow as 1 ns. 
Sample - Sample data only. 
Envelope - Max/Min values acquired over one or more acquisitions. 
Average - Waveform data from 2 to 572 (selectable) acquisitions is averaged. 
Single Sequence - Use the Single Sequence button to capture a single triggered acquisition sequence at a time.

Trigger System

Main Trigger Modes - Auto (supports Roll Mode for 40 ms/div and slower), Normal.
B Trigger - Trigger after time or events.
 
Trigger After Time Range - 13.2 ns to 50 s. 
 
Trigger After Events Range - 1 to 9,999,999 events. 
 
External Trigger Input  - >1 megaohm in parallel with 17 pF; Max input voltage is 150 VRMS.

Trigger Types

Edge - Conventional level-driven trigger. Positive or negative slope on any channel. Coupling selections: DC, noise reject, HF reject, LF reject.
 
Video - Trigger on all lines or individual line, odd/even or all fields or analog HDTV formats (1080i, 1080p, 720p, 480p). See optional TDS3VID and TDS3SDI application modules for extended video triggering and measurement features. 
 
Logic  - (requires TDS3TRG)
PATTERN: Specifies AND, OR, NAND, NOR when true or false for a specific time. 
STATE: Any logic state. Triggerable on rising or falling edge of a clock. 

Note: Logic triggers can only be used on combinations of 2 inputs. 

Pulse  -  (requires TDS3TRG)
WIDTH (or GLITCH): Trigger on pulse width less than, greater than, equal to or not equal to a selectable time limit ranging from 39.6 ns to 50s. 

RUNT: Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again. 

SLEW RATE: Trigger on pulse edge rates that are either faster or slower than a set rate. Edges can be rising, falling or either. 

Comm (requires TDS3TMT) - provides isolated pulse triggering required to perform DS1/DS3 telecommunications mask testing per ANSI T1.102 standard. 

Alternate - Sequentially uses each active channel as a trigger source.

Measurement System

Automatic Waveform Measurements - Period, Frequency, +Width, -Width, Rise-time, Fall-time, +Duty Cycle, -Duty Cycle, +Overshoot, -Overshoot, High, Low, Max, Min, P-P, Amplitude, Mean, Cycle Mean, RMS, Cycle RMS, Burst Width. 

Display any four measurements from any combination of waveforms. 

Thresholds - Settable in percentage or voltage. 

Gating - Measurements can be gated using the screen or vertical cursors.

Waveform Processing

Deskew - Channel-to-Channel deskew ±10 ns may be manually entered for better timing measurements and more accurate math waveforms.
 
Arithmetic Operators - Add, subtract, multiply, divide. 
 
Autoset - Single-button, automatic setup on selected input signal for vertical, horizontal and trigger systems.

Display Characteristics

Waveform Style - Dots, vectors and variable persistence.
 
Graticules - Full, grid, cross-hair, frame, NTSC, PAL, SECAM, vectorscope 100% and 75% color bars (with optional TDS3VID and TDS3SDI video application modules). 
 
Format - YT, XY and Gated XYZ (XY with Z-axis blanking available on TDS30X4 only).

I/O Interface

Hardcopy Port (standard) - Centronics-type parallel.
 
TDS3GM Communications Module - 

GPIB (IEEE -488.2) Programmability: Full talk/listen modes; Control of all modes, settings and measurements. 
RS-232-C Interface Programmability: Full talk/listen modes; Control of all modes, settings and measurements. Baud Rate up to 38,400. DB-9 male connector. 
Programmer Manual: 071-0381-01. 

TDS3VM Communications Module - 

VGA: Monitor output for direct display on large VGA-equipped monitors. DB-15 female connector, 31.6 kHz sync rate, EIA RS-343A compliant. 
RS-232-C Interface Programmability: same as TDS3GM. 
Programmer Manual: same as TDS3GM. 

TDS3EM Communications Module - 

Ethernet Port: 10Base-T with RJ-45 connector. Provides local area network printing and programming interface. 
RS-232-C Interface programmability: same as TDS3GM. 
Programmer Manual: same as TDS3GM. 

Ethernet Built-In Communications Module - Ethernet Port: 10Base-T with RJ-45 Connector. 
Provides local area network printing and programming interface. 
RS-232-C Interface programmability: same as TDS3GM. 
Programmer Manual: Same as TDS3GM. 

Note: Only one Communications Module may be installed at a time. 
All Communication Modules include WaveStarTM Software for oscilloscopes 30-day, full-functioning product demo.

Hard Copy Capability

Graphics File Formats - Interleaf (.img), TIF, PCX (PC Paintbrush), BMP (Microsoft Windows) and Encapsulated Postscript (EPS).
 
Printer Formats - Bubblejet, DPU-3445, Thinkjet, Deskjet, Laserjet, Epson (9- and 24-Pin).

Environmental and Safety

Temperature - +5 to +50°C (operating), -20 to +60°C (nonoperating).
 
Humidity - 20% to 80% RH below 32°C, derate to 30% RH at 45°C (operating), 5% to 90% RH below 41°C, derate to 30% RH at 60°C (nonoperating). 
 
Altitude - to 3,000 m (operating), 15,000 m (nonoperating). 
 
Electromagnetic Compatibility - Meets or exceeds EN55011 Class A Radiated and Conducted Emissions; EN50082-1; FCC 47 CFR, Part 15, Subpart B, Class A; Australian EMC Framework; Russian GOST EMC regulations. 
 
Safety - UL3111-1, CSA1010.1, EN61010-1, IEC61010-1.

Physical Characteristics

Instrument
Dimensions mm in.
Width 375.0 14.8
Height 176.0 6.9
Depth 149.0 5.9
Weight kg lbs.
Instrument only 3.2 7.0
  w/battery 5.2 11.5
Instrument Shipping Package
Dimensions mm in.
Width 502.0 19.8
Height 375.0 14.8
Depth 369.0 14.5
Rackmount
Dimensions mm in.
Width 484.0 19.0
Height 178.0 7.0
Depth 152.0 6.0