Signet Indicators and Controllers
Signet Intelek-Pro™ Family
- Micro-processor based controllers
- Modular plug-in input/output cards
- Front-panel access to electronics
- Operating Temperature: 32 to 130°F (0 to 55°C)
- Power Requirements: 17 to 30 VDC @ 0.5 Amps max.; 90 to 132 VAC; 180 to
264 VAC
- ABS Plastic enclosure, NEMA 4X (IP65) front/rear cover (optional)
- Dimensions: 3.5" x 3.5" x 6.5"; (88 x 88 x 165 mm)
- LCD: 4.5 digits, 0.5" height (upper) 8 digits, 0.3" height (lower)
9010 Flow Controller --
- Security access code capability
- Compact NEMA 4X/IP65 rated 1/4 DIN enclosure
- Optional analog input
- Engineered for liquid flow measurement and control in process pipes
- Input Frequency Range: 0 to 10 KHz
- Optional Inputs: 4 to 20 mA or 0 to 20 mA/0 to 5 VDC or 0 to 10 VDC
- Isolation: 500 VDC to earth ground
- 2-Relay Output Card: 2 SPDT contact outputs, 5 A @ 250 VAC or 30 VDC resistive
load
- Display Accuracy: 0 to 5 KHz input, ±0.05% of reading
9030 pH Controller --
- Isolated Inputs to 500 VDC
- Up to four alarms
- Modular " plug-in" output cards
- AC/DC power capability
- Modular design allows operator to add control option cards as system requirements
increase
- Output options include up to four 5 amp relays, for on/off alarm or proportional
pulse operation, and one or two analog outputs
- pH Range: 0.00 to 14.00 pH
- Display Accuracy: ±0.02 pH
- Analog Output Accuracy: ±0.5% fs
- Temp. Input Range: -13 to 221°F
- Maximum Output Configuration: 1 relay card & 2 analog outputs, or 1
relay card, 1 proportional card, and 1 analog output
9040 ORP Controller --
- Raw mV display
- Input isolation
- mV Range: ±3000 mV (Xl) gain ±1000 mV (X-3) gain
- Display Accuracy: ±0.1% of f.s.
- Analog Output Accuracy: ±0.5% of f.s.
- Ambient Operating Temperature: 32 to 130°F (0 to 55°C)
- Input Isolation: 500 VDC to earth ground
9050 Dual Channel Conductivity --
Controller
- Control of conductivity, temperature or percent rejection
- Modular " plug-in" output cards
- Dual channel inputs
- AC/DC power capability
- Modular design allows operator to add control option cards as system requirements
increase
- Output options include up to four 5 amp relays, for on/off alarm or proportional
pulse operation, and one or two analog outputs
- Accepts input from up to two sensors with varying ranges, for independent
control of each channel, as well as percent reject output for water treatment
applications
- Conductivity range: 1 to 200,000 µS, 0-10 M(omega)
- Temperature range: 0 to 100°C
5500 Flow Monitor --
- 12 to 24VDC/AC powered
- Independently calibrated flow and totalizer displays
- Quality standards: UL, CSA, CE
- Two 8-digit totalizers (resettable and non-resettable)
- Remote totalizer reset from front keypad or by external contact closure
- Current, frequency and alarm outputs
- 6 dial ranges with assorted flow rate/totalizer decals included
- Non-volatile memory
- Splashproof front panel
- Scaleable (totalizer) pulse output
- Accuracy: ±0.5% of reading
- Repeatability: ±0.2% of reading
- Input: optically isolated
- Input range: 2Hz to 10 kHz
5090 Flow Monitor --
- Sensor powered, requires no external power source
- Simple front panel - 1-point calibration
- Quality standards: FM, UL, CSA, CE
- Factory Mutual (FM) approved for intrinsic safety in classes I, II, III
division 1 hazardous locations without intrinsic safety barriers
- Splashproof, shatter resistant acrylic front panel
- 6 dial ranges with assorted flow rate/multiplier decals included
- Wide sensor range: from 1 to 20 fps in 0.5 to 36" pipes
- 1/4 DIN panel cutout
- Accuracy: ±2% of full scale
- Repeatability: ±1% of full scale
2550 Insertion Magmeter --
- Insertion technology easy to install and remove
- 50/60 Hz noise rejection
- Fluid diagnostics
- Flow Range: 0.3 to 20 ft/s
- Adjustable to various pipe sizes
- Maximum Operating Pressure: 250 PSIG/17 bar
- Fluid Temperature: 32 to 212°F
- Power Requirements: 24 VDC ± 2.4 VDC
- Outputs: 1--Isolated 0 to 20 or 4 to 20 mA; 2--Isolated open collector
- Repeatability: 0.5% of full scale
5700 pH/ORP Monitor --
- 12 to 24VDC/AC powered
- pH or ORP operation with auto sensor recognition
- EASY CAL feature with auto buffer recognition
- Quality standards: UL, CSA, CE
- Current and alarm outputs
- Reversible pH and ORP dial included
- Non-volatile memory
- Splashproof front panel
- Accuracy: ±0.2% of reading
- Simple push-button operation
- Auto or manual calibration menu
5800 Conductivity/Resistivity Monitor --
- 12 to 24VDC/AC powered
- Quality standards: UL, CSA, CE
- Display units: µS, mS, k(omega), M(omega), PPM
- Current and alarm outputs
- 6 dial ranges with assorted conductivity unit/multiplier decals included
- Non-volatile memory
- Splashproof front panel
- Accuracy: ±2% of reading
- Simple push-button operation
- Programmable temperature compensation and TDS conversion factors
5100 Flow Monitors --
- Power: internal battery
- Independently calibrated flow and totalizer displays
- Quality standards: UL, CSA, CE
- 4-digit flow display
- Two 8-digit totalizers (resettable and non-resettable)
- 5100: remote totalizer reset from front keypad or by external contact closure
- Splashproof front panel
- Socketed battery for easy replacement through rear access panel
- Assorted flow rate/totalizer unit decals included
- Accuracy: ±0.5% of reading
|