Hart Scientific Temperature
Calibration Equipment
Hart Scientific 9009 Industrial Dual-Block Calibrator
 |
Feature Summary
- Temperatures from –15°C to 350°C in one unit
- Two wells in each block for simultaneous comparison calibrations
- Rugged, lightweight, water-resistant enclosure
- Controlled by precision Hart Scientific temperature controller
Ordering Info
|
The Model 9009 includes two independently controlled temperature blocks. The hot block provides temperatures from 50°C to 350°C, while the cold block covers the range –15°C to 110°C. Each block is controlled by a precision Hart Scientific temperature controller. These aren’t some off-the-shelf controllers we glued into a box. These are Hart Scientific controllers from the leading temperature company in the world.
Each temperature block includes two wells with removable inserts. You can calibrate four probes at once, or you can calibrate two probes at the same time with an external reference (like Hart’s Model 1521 LLK Thermometer), or you can use the two temperature wells to get quick “zero” and “span” references for transmitter calibrations.
The 9009 is housed in a tough Pelican™ case that is both airtight and watertight. It’s a small package weighing only 10 pounds, yet it fits everything you need, including a power cord and four extra inserts. Inserts are available to accommodate sensors of any size from 1/16" (1.6 mm) to 7/16” (11.1 mm). This rugged system can go anywhere.
| Specifications |
Hot Block
|
Cold Block
|
| Range |
50°C to 350°C (122°F to 662°F)
|
–15°C to 110°C (5°F to 230°F)
(–8°C [18°F] with hot block at 350°C [662°F])
|
| Accuracy |
±0.6°C
|
±0.2°C
|
| Stability |
±0.05°C
|
| Well-to-Well Uniformity |
±0.1°C
|
| Display Resolution |
0.1°
|
| Heating Times |
10 minutes from 25°C to 350°C
|
15 minutes from 25°C to 110°C
|
| Cooling Times |
30 minutes from 350°C to 100°C
|
16 minutes from 25°C to –15°C
|
| Stabilization Times |
8 minutes
|
| Well Depth |
4" (102 mm)
|
| Removable Inserts |
Two 1/4” (6.4 mm) and 3/16” (4.8 mm) inserts included; see Ordering Information for other available inserts
|
| Computer Interface |
RS-232 included with free Interface-it software
|
| Power |
115 VAC (±10%), 3 A, or
230 VAC (±10%), 2 A, specify, 50/60 Hz, 250 W
|
| Size |
7" H x 10.5" W x 9.75" D (178 x 267 x 248 mm)
|
| Weight |
10 lb. (4.5 kg)
|
| NIST-Traceable Calibration |
Data at 50°C, 100°C, 150°C, 200°C, 250°C, 300°C, and 350°C
|
Data at –8°C, 0°C, 25°C, 50°C, 75°C, 100°C, and 110°C
|
Hart Scientific Field Dry Well Calibrators
 |
Models 9103, 9140, and 9141
- RS-232 and Interface-it software included
|
These three units beat every other comparable dry-well in the industry in performance, size, weight, convenience, ease of calibration, software, and price. In addition, the heating and cooling rate of each of these dry-wells is adjustable from the front panel, thermal switches can be checked for actuation testing, and multiple-hole inserts are available for a variety of probe sizes.
Hart dry-wells are easy to calibrate. You don’t even have to open the case. This means less maintenance costs and less down time when they do need calibration.
Our Interface-it software lets you adjust set-points and ramp rates, log dry-well readings to a file, create an electronic strip chart, and perform thermal switch testing with data collection. The software is written for Windows and has a great graphical interface. It’s the best dry-well software in the industry. Regardless of whether you want basic software or a completely automated calibration system, we’ve got what you want.
| Specifications |
9103
|
9140
|
9141
|
| Range |
–25°C to 140°C (–13°F to 284°F)
at 23°C ambient
|
35°C to 350°C (95°F to 662°F)
|
50°C to 650°C (122°F to 1202°F)
|
| Accuracy |
±0.25°C
|
±0.5°C (holes greater than 1/4" [6.35 mm]: ±1°C)
|
±0.5°C to 400°C; ±1.0°C to 650°C
(holes greater than 1/4": ±2°C)
|
| Stability |
±0.02°C at –25°C
±0.04°C at 140°C
|
±0.03°C at 50°C
±0.05°C at 350°C
|
±0.05°C at 100°C
±0.10°C at 500°C
±0.12°C at 650°C
|
| Well-to-Well Uniformity |
±0.1°C between similarly sized wells
|
±0.1°C with similarly sized wells
|
±0.1°C below 400°C, ±0.5°C above 400°C with similarly sized wells
|
| Heating Times |
18 minutes from ambient to 140°C
|
12 minutes from ambient to 350°C
|
12 minutes from ambient to 650°C
|
| Cooling Times |
20 minutes from ambient to –25°C
|
15 minutes from 350°C to 100°C
|
25 minutes from 650°C to 100°C
|
| Stabilization Time |
7 minutes
|
| Immersion Depth |
4.875” (124 mm)
|
| Inserts |
Insert A, B, or C included (specify when ordering)
|
| Computer Interface |
RS-232 included with free Interface-it software (Model 9930)
|
| Power |
115 VAC (±10%), 1.3 A or 230 VAC (±10%), 0.7 A, switchable, 50/60 Hz, 150 W
|
115 VAC (±10%), 4.4 A or 230 VAC (±10%), 2.2 A, switchable, 50/60 Hz, 500 W
|
115 VAC (±10%), 8.8 A or 230 VAC (±10%), 4.4 A, switchable, 50/60 Hz, 1000 W
|
| Size |
5.63" W x 10.25" H x 9.63" D
(143 x 261 x 245 mm)
|
6" W x 3.375" H x 7.75" D
(152 x 86 x 197 mm)
|
4.3" W x 9.3" H x 7.3" D
(109 x 236 x 185 mm)
|
| Weight |
12 lb. (5.7 kg)
|
6 lb. (2.7 kg)
|
8 lb. (3.6 kg)
|
| NIST-Traceable Certificate |
Data at –25°C, 0°C, 25°C, 50°C, 75°C, 100°C, and 140°
|
Data at 50°C, 100°C, 150°C, 200°C, 250°C, 300°C, and 350°C
|
Data at 100°C, 200°C, 300°C, 400°C, 500°C, and 600°C
|
Hart Scientific High Accuracy Dry-Wells
Model 9105 Low Temp includes 1/4" insert
Model 9107 Ultra Low Temp includes 1/4" insert
Model 9122 High Capacity includes 1/8", 3/16", 3/8", and 1/4" inserts
Model 9107 High Speed includes multi-hole insert (specify)
Features:
- World’s best dry-wells—accurate to 0.1°C, stable to ±0.005°C
- Largest-capacity temperature wells
- Model 9107 reaches lowest temperatures: –45°C in 23°C ambient
- Model 9122A provides temperatures to 700°C
| Specifications |
9105 |
9107 |
9122A |
9127 |
| Range |
–25°C to 140°C
(–13°F to 284°F)
at 23°C ambient |
–45°C to 140°C
(–49°F to 284°F)
at 23°C ambient |
50°C to 700°C†
(122°F to 1292°F) |
50°C to 600°C
(122°F to 1112°F) |
| Accuracy |
Center well:
±0.1°C |
Center well:
±0.1°C |
Center well:
±0.1°C at 100°C
±0.1°C at 300°C
±0.3°C at 660°C |
±0.1°C at 100°C
±0.15°C at 300°C
±0.5°C at 600°C
|
| Stability |
±0.01°C |
±0.01°C at –40°C
±0.005°C at 0°C
±0.005°C at 100°C |
±0.01°C at 100°C
±0.02°C at 300°C
±0.05°C at 660°C |
±0.01°C to 100°C
±0.02°C to 300°C
±0.05°C to 600°C |
| Well-to-Well Uniformity |
Drilled wells:
±0.05°C |
Drilled wells:
±0.05°C |
Drilled wells:
±0.025°C at 100°C
±0.1°C at 300°C
±0.3°C at 660°C |
±0.05°C |
| Well Depth |
6" (152 mm)
|
| Computer Interface |
RS-232 Interface included with Model 9930 Interface-it control software, IEEE optional |
| Heating Time to Max. |
10 minutes |
15 minutes |
75 minutes |
30 minutes |
| Cooling Time |
25°C to –25°C: 15 minutes |
25°C to –45°C: 35 minutes |
700°C to 100°C: 280 minutes |
600°C to 100°C: 125 minutes |
| Test Wells |
5 wells: 2 at 1/4" (6.35 mm), 1 at 3/16" (4.8 mm), 1 at 1/8" (3.2 mm), and 1 interchangeable (3/4"/19.1 mm O.D.) |
5 wells: 2 at 1/4" (6.35 mm), 1 at 3/16" (4.8 mm), 1 at 1/8" (3.2 mm), and 1 interchangeable (3/4"/19.1 mm O.D.) |
9 wells: 4 interchangeable (3/4"/19.1 mm O.D.) and 5 at 1/4" (6.35 mm) |
1 interchangeable well accommodates multi-hole insert (1.5"/38.1 mm O.D.) |
| Resolution |
0.01°C or °F |
| Display |
LED, °C or °F, user-selectable |
| Size |
13.5" H x 7.8" W x 11.9" D (343 x 198 x 302 mm) |
| Weight |
26 lb. (11.8 kg) |
22 lb. (10 kg) |
25 lb. (11.3 kg) |
25 lb. (11.3 kg) |
| Power |
115 VAC (±10%), 3 A or 230 VAC (±10%), 1.6 A, specify, 50/60 Hz, 350 W |
115 VAC (±10%), 4 A or 230 VAC (±10%), 3.15 A, specify, 50/60 Hz, 350 W |
115 VAC (±10%), 8.8 A or 230 VAC (±10%), 4.4 A, specify, 50/60 Hz, 1000 W |
115 VAC (±10%), 8.8 A or 230 VAC (±10%), 4.4 A, switchable, 50/60 Hz, 1000 W |
| NIST-Traceable Calibration |
Data at –25°C, 0°C, 75°C, and 140°C |
Data at –45°C, 0°C, 75°C, and 140°C |
Data at 100°C, 200°C, 300°C, 400°C, 500°C, and 660°C |
Data at 100°C, 200°C, 300°C, 400°C, 500°C, and 600°C |
| †Calibrated to 660°C; reference thermometer recommended at higher temperatures. |
Hart Scientific Calibration Furnace --- 1200 °C
 |
Thermocouple Furnace Model 9150
- Low-cost thermocouple furnace
- NIST-traceable calibration included
- RS-232 port standard
With a stability of ±0.5°C, it has a temperature range to 1200°C and a display accuracy of ±5°C across its entire range.
Ordering Info
|
With interchangeable temperature blocks, you can check thermocouples as small as 1/16 of an inch in diameter. The 9150 works with 115 or 230 VAC power.
|
 |
| Specifications |
| Temperature Range |
150°C to 1200°C
(302°F to 2192°F) |
| Display Resolution |
0.1° to 999.9°
1° above 1000° |
| Stability |
±0.5°C |
| Display Accuracy |
±5°C |
| Well Diameter |
1.25" (32 mm) |
| Well Depth |
4" (102 mm) |
| Heating Time |
35 minutes to 1200°C |
| Cooling Time |
140 minutes with block |
| Well-to-Well Uniformity |
±0.5°C to ±1.0°C
(Insert “C” at 1200°C) |
| Stabilization |
20 minutes |
| Power |
115 VAC (±10%), 10.5 A or 230 VAC (±10%), 5.2 A, switchable, 50/60 Hz, 1200 W |
| Size |
12.4" H x 8.2" W x 12.4" D
(315 x 208 x 315 mm) |
| Weight |
28 lb. (13 kg) |
| NIST-Traceable Calibration |
Data at 150°C, 300°C, 450°C, 600°C, 800°C, 1000°C, 1200°C |
Portable Infrared Calibrators
 |
Features
-
Certify IR pyrometers from 30°C to 500°C
(22°F to 932°F)
-
Large 2.25" (57 mm) blackbody target
-
RTD reference well for high precision
-
Small, compact design
Ordering Info |
These new units feature
a large, temperature controlled blackbody target with a diameter of 2.25"
(57 mm), which offers a large field of view area for optical variations
in infrared thermometers. The emissivity of the isothermal target
is set at 0.95, and the target temperature can be controlled in set-point
increments of 0.1° from 30°C to 500°C.
For even higher precision,
a well is located directly behind the blackbody surface for contact calibration
of the blackbody. Using an optional digital RTD thermometer, target
accuracy of ±0.075°C can be achieved.
These units are as
easy to use as point-and-shoot. Simply set the desired blackbody
temperature from the convenient front panel control buttons, wait a few
minutes for equilibrium, and point the gun at the target. The radiated
energy from the blackbody is measured by your IR thermometer. Simply
compare its reading to the display on the blackbody and record the difference.
| Specifications |
9132
|
9133
|
| Temperature
Range |
50°C to 500°C
(122°F to 932°F)
|
30°C to 150°C at 23°C
ambient
(22°F to 302°F at 73°F ambient)
|
| Accuracy |
±0.5°C at 100°C (±0.9°F
at 212°F)
±0.8°C at 500°C (±1.4°F
at 932°F)
|
±0.4°C (±0.72°F)
|
| Stability |
±0.1°C at 100°C (±0.18°F
at 212°F)
±0.3°C at 500°C (±0.54°F
at 932°F)
|
±0.1°C (±0.18°F)
|
| Target
Size |
2.25" (57 mm)
|
| Target
Emissivity |
0.95
|
| Resolution |
0.1°
|
| Heating
Time |
30 minutes (50°C to 500°C)
|
15 minutes (25°C to 150°C)
|
| Cooling
Time |
30 minutes (500°C to 100°C)
|
15 minutes (25°C to 20°C)
|
| Computer
Interface |
RS-232 included with 9930 Interface-it
software
|
| Power |
115 VAC (±10%), 3 A or 230
VAC (±10%), 1.5 A, switchable, 50/60 Hz
|
115 VAC (±10%), 1.5 A, 230
VAC (±10%), 1.0 A, switchable, 50/60 Hz
|
| Size |
4" H x 6" W x 7" D (102 x 152 x 178
mm)
|
6" H x 11.25" W x 10.5" D (152 x 286
x 267 mm)
|
| Weight |
4 lb. (1.8 kg)
|
10 lb. (4.6 kg)
|
|