Contact the closest office: 1-800-358-5525

  Hart Scientific Precision Temperature Calibrator / Data Logger

Model 1560 Black Stack 

Buy It Now ! Ordering Info

  • Reads SPRTs, RTDs, thermistors, and thermocouples
  • Any configuration you like up to eight modules
  • High-accuracy reference thermometer (to ±0.0013°C)
  • Automates precision data acquisition
  • LCD Screen capable of displaying data in graphical strip-chart format
  • RS232 port standard, IEEE488 port is optional
The 1560 Black Stack can be any kind of thermometer you want it to be, and it works in three distinctive ways.

It’s a reference thermometer with a  NIST traceable calibration; it’s an automated calibration system reading your reference probe and sensors you’re testing; or it’s a high-accuracy data acquisition system.  And it does each of these functions better than any other thermometer currently on the market.

The Stack consists of up to eight different modules that fit together to do any type of thermometry you choose.  You can buy all of them, or any combination of them, and change the Stack and its functions anytime you want. Each module stacks behind the preceding one, and when you add a module, the Stack’s software automatically reconfigures itself to include all of the new functions supplied by that module.  There’s nothing to take apart.  No boards need to be installed.  There’s no software to load, and nothing has to be calibrated. 

Hart’s new Model 9935 Multi Channel LogWare makes the Black Stack an even more powerful data acquisition tool. LogWare provides graphical and statistical analysis of each channel you’re measuring (up to 96 with the Black Stack). And with alarms that can be customized, delayed start times, and selectable logging intervals, LogWare turns the Black Stack into the most powerful temperature data acquisition tool on the market.
 
Specifications
Model 1560 Base Unit
Power: 100 to 240 VAC, 50 or 60 Hz, nominal; Attachable Modules: up to 8; Display: 4.25" x 2.25" LCD graphics, LED backlight, adjustable contrast and brightness; Automatic Input Sequencing: 1 to 96 channels; Communications: RS-232; Non-volatile Memory: channel sequence, probe coefficients; Minimum Sample Time: 2 seconds.
Extended Communication Module 3560
The Extended Communication Module adds additional communication interface capability to the system. This module includes a GPIB (IEEE-488) interface, Centronics printer interface, and analog output. The GPIB interface connects the 1560 to a GPIB bus. GPIB can be used to control any function of the 1560 and read measurement data. The printer interface allows the 1560 to send measurement data directly to a printer. The analog output sources a DC signal (±1.25 VDC) corresponding to the value of a measurement.
Resistance Modules
Input 
Channels
Resistance Range
Basic
Resistance Accuracy
Resistance Resolution
Temperature Range
Equivalent Temperature Accuracy
Temperature Resolution
SPRT Module 2560
2
0W to 400W
±20 ppm of reading (0.0005W at 25W
0.002W at 100W)
0.0001W
–260°C to 962°C
±0.005°C at 0°C
±0.007°C at 100°C
0.0001°C
High-Temp PRT Module 2561
2
0W to 25W
±50 ppm of reading 
(0.00013W at 2.5W)
0.00001W
0°C to 1200°C
±0.013°C at 0°C
±0.018°C at 100°C
0.001°C
PRT Scanner 2562
8
0W to 400W
±40 ppm of reading 
(0.004W at 100W)
0.0001W
–200°C to 850°C
±0.01°C at 0°C
±0.014°C at 100°C
0.0001°C
Standards Thermistor Module 2563
2
0W to 1 MW
±50 ppm of reading 
(0.5W at 10 KW)
0.1W
–60°C to 260°C
±0.0013°C at 0°C
±0.0015°C at 75°C
0.0001°C
Thermistor Scanner 2564
8
0W to 1 MW
±100 ppm of reading 
(1W at 10 KW)
0.1W
–60°C to 260°C
±0.0025°C at 0°C
±0.003°C at 75°C
0.0001°C
1000W PRT Module 2567
2
0W to 4 KW
±25 ppm of reading
(0.025 at 1 KW)
0.001W
–260°C to 962°C
±0.006°C at 0°C
±0.009°C at 100°C
0.0001°C
1000W PRT Scanner 2568
8
0W to 4 KW
±40 ppm of reading
(0.04 at 1 KW)
0.001W
–200°C to 850°C
±0.01°C at 0°C
±0.014°C at 100°C
0.0001°C
Thermocouple Modules
 

Input 
Channels
 

Millivolt Range

 

Millivolt Accuracy

 

Millivolt Resolution
Temperature Accuracy,
Ext. CJC
Temperature Accuracy,
Int. CJC
 

Temperature Resolution
Precision Thermocouple Module 2565
2
–10 to 100 mV
±0.002 mV
0.0001 mV
±0.05°C
±0.1°C
0.001°C
Thermocouple Scanner 2566
12
–10 to 100 mV
±0.004 mV
0.0001 mV
±0.1°C
±0.3°C
0.001°C
†  Temperature accuracy depends on probe type and temperature.

 


About...  |  Table of Contents  | Manufacturer Index  | What's New

Send Inquiries to:

Copyright © 2006 Davis Inotek Instruments, LLC. All rights reserved.